MED-EL OPUS2
List of required parts
1
OPUS2 speech processor
2
FM Battery Pack Cover
3
MLxi or MicroMLxS
Fitting with the user
1. Turn all equipment Off.
2. Replace the standard battery cover by the FM Battery Pack
Cover.
3. Plug the FM receiver into the bottom of the FM Battery
Pack Cover. The MicroMLxS is usually used in the two green
dot position.
4. Switch the speech processor and FM transmitter On.
5. Test the patient’s speech recognition in quiet with the
CI alone and then through the wireless FM system only
(out of range of the processor microphone). Performance
should be similar between these two conditions. If the
FM is too loud or too soft, the FM programming software
from Phonak may be required to optimize the FM volume
level.
6. Evaluate FM benefit as well as the appropriateness of FM
settings trough speech-recognition assessment in the
clinic and standardized questionnaires completed by the
recipient, family members and teachers.
Additional information
When MicroMLxS is used, the switch acts as following:
a. Two green dots: FM + M
b. One green dot: FM only (processor microphones
attenuated)
When MLxi is used and the push button is enabled, the push
button acts as following:
a. Long press: switch MLxi On or Off respectively
b. Short press: switch between FM + M and FM only in case
a non Dynamic FM transmitter is being used.
Listening check
Plug the MLxi or MicroMLxS into the headset checker (See
page 12). Turn On the headset checker and FM transmit-
ter and you should hear the FM signal.
Plug the MLxi or MicroMLxS into an audioshoe attached
to a hearing instrument. Turn On the hearing instrument
and switch it to the FM program if required.